A Neuro-Inspired Chip for Constant Heart Rate Monitoring

Smartwatches and other wearables are transforming health monitoring, offering accurate heart rate tracking and anomaly detection. These devices, powered by advanced sensors and machine learning, analyze biosignals – the body’s electrical and optical measurements. However, their small size and battery constraints limit continuous, long-term monitoring. To overcome these limitations, researchers Chiara De Luca, Mirco Tincani, Giacomo Indiveri, and Elisa Donati explored neuromorphic technology, mimicking the human brain’s energy-efficient processing. This innovative approach promises to enable wearable gadgets that provide uninterrupted health monitoring with minimal maintenance.
The researchers developed a system that integrates multiple biosignals, such as heartbeats and movement, using neural computation techniques. This system, tested on a mixed-signal neuromorphic processor, demonstrates remarkable resilience to the inherent variability of analog circuits. It efficiently processes heart rate data, providing real-time beat-per-minute readings, categorizes heart rate zones, and tracks long-term trends indicative of stress or agitation. Such devices could operate autonomously for extended periods, offering unobtrusive, continuous health monitoring.
Our brains effortlessly process a constant stream of sensory data with exceptional energy efficiency. This new system seeks to replicate this by using analog neuromorphic circuits, which function with continuous signals akin to brain activity. Despite inherent variations, these circuits offer significant power savings and handle complex computations. By employing artificial neuron groups, the system achieves robust and reliable biosignal interpretation, even amidst noise and hardware inconsistencies.
This neuromorphic approach holds immense potential in health technology. Similar systems have successfully identified anomalies in ECG, EMG, and EEG signals from epilepsy patients. The researchers configured their artificial neurons to fuse data from diverse sources and analyze it across various time scales. This capability could prove invaluable for monitoring heart health in elderly individuals or those with memory impairments, where continuous monitoring is crucial yet device management can be challenging.
The system’s input comprises biosignals from photoplethysmography (PPG), electrocardiography (ECG), and accelerometers. These signals undergo bandpass filtering to isolate heart rate-related patterns. When motion artifacts contaminate the data, a Nelder-Mead algorithm blends motion and heart signals for noise reduction. Each filtered signal then activates a Leaky Integrate-and-Fire neuron, generating spikes that represent the processed data.
The spikes are then processed through two pathways. One pathway feeds a decoder, providing real-time heart rate measurements. The other pathway employs a Neural State Machine (NSM) to track heart rate zone transitions. The decoder utilizes artificial neuron groups to correlate spike patterns with heart rate values. The NSM, inspired by animal brain mechanisms, averages activity across neurons and balances excitation and inhibition for robust zone tracking.
The system was implemented on a Dynamic Neuromorphic Asynchronous Processor (DYNAP-SE) chip and tested over several days. The chip exhibited consistent performance, demonstrating resilience to hardware variations. Three NSM configurations were evaluated for heart rate zone handling. The first, a soft Winner-Take-All network, quickly identifies the dominant heart rate zone. The second, a Nearest Neighbors NSM, smooths transitions and reduces noise. The third, a Monotonic NSM, tracks gradual heart rate increases, useful for stress or exertion monitoring.
Using clean hospital data, the system accurately identified heart rate zones. Wrist-worn data, more susceptible to motion artifacts, required signal blending with accelerometer data. The decoder, enhanced with Gaussian-weighted distributions, achieved heart rate accuracy within one beat per minute, even with overlapping filter zones. This overlap enhanced accuracy and reduced sensitivity to minor fluctuations.
Power consumption tests revealed remarkable efficiency. The chip consumed only 13.1 microwatts in a five-second test, translating to days of operation on a single charge. The system demonstrated robust performance with both synthetic and real-world data, adapting to signal strength and noise variations. The ‘nearby zone’ configuration excelled at smoothing transitions, particularly at extreme values, ensuring low error rates.
The Monotonic NSM configuration, designed for tracking gradual heart rate increases, proved effective for monitoring exertion and detecting stress-related trends. Tested with wrist-worn data from various activities, the system demonstrated robust noise reduction. Optimized signals, generated using a Nelder-Mead algorithm, accurately tracked heart rate zones, highlighting the system’s adaptability.
A key advantage lies in the system’s on-chip processing, eliminating the need for power-intensive software. Mixed-signal neuromorphic circuits perform signal extraction and noise reduction at the hardware level. The chip’s filters, capable of expansion, achieved heart rate accuracy within one beat per minute. The system’s ability to handle short and long-term monitoring on a single low-power chip makes it ideal for continuous, unobtrusive wearable applications.
In summary, exercise-specific tuning achieves comparable accuracy to personalized calibration, significantly enhancing the system’s real-world applicability. This streamlined methodology outperforms complex alternatives like data intensive deep learning and computationally expensive offline methods. Utilizing E-I balanced networks and Neural State Machines, the system achieves robust stability and smooth transitions, minimizing noise and power consumption. The Winner-Take-All approach ensures clear heart rate zone identification. Crucially, this system’s ability to provide continuous, accurate, and low-power health monitoring holds the potential to save lives. By detecting subtle, long term changes in heart rate and activity, it can provide early warnings of critical health issues, enabling timely intervention and improving patient outcomes. This technology represents a significant step towards proactive healthcare, where wearable devices are not just tools for fitness tracking, but vital instruments for life saving health monitoring.
Citation:
De Luca, C., Tincani, M., Indiveri, G., & Donati, E. (2025). A neuromorphic multi-scale approach for real-time heart rate and state detection. npj Unconventional Computing, 2, Article number: 6. Published by Nature. Available at https://www.nature.com/articles/s44335-025-00024-6
NeuromorphicCore